Ultrasound bath ВУ-09-"Я-ФП"-02

All parts of the bath (body, working container, cover) are made of stainless steel.
All parts of the bath (body, working container, cover) are made of stainless steel.

A piezoelectric element is installed at the bottom of the bath, which converts ultrasonic energy supplied from the generator into mechanical vibrations of the bath bottom and walls leading to cavitation in the washing solution.

The bath is controlled by built-in electromechanical timer with maximum cleaning interval of 120 minutes. Manual activation method of constant operation mode is variably permitted.

An indicator on the front panel informs that cleaning mode is on.
MAIN PERFORMANCE VALUES

Ultrasonic transducer resonance frequency, kHz, - 40±2.
Treatment procedure timer, min. . . . 10-120.
(for models ВУ-09-"Я-ФП"- 02).
Supply voltage, V. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22010%, 50/60 Hz
The bath overall dimensions are 175х160х225 mm*
* Permissible deviation, no more than ± 5 mm.
The bath working container overall dimensions are 150х140х100 mm*
* Permissible deviation, no more than ± 5 mm.
The bath working container volume - 1.7 l*
* Permissible deviation, no more than ± 0.1 l
The bath working container useful volume - 1.2 l*
* Permissible deviation, no more than ± 0.1 l.
Bath weight (outfitted) - 2.6 l*
* Permissible deviation, no more than +/- 0.5 kg.
Power consumed by the bath - 200 VA
The liquid cavitation is provided in the bath over the entire useful volume of the working container. The cavitation effect is provided by the ultrasonic transducer.
The bath is equipped with a treatment procedure timer providing for control of instruments treatment time.
Timer operation parameters:
Range 120 min
Increment 10 min
Permissible deviation ± 4
The bath has a metal mesh located in the working container with the help of holders to accommodate the instruments withstanding uniformly-distributed load of 40 N
The time to recover the bath operation mode after switching on does not exceed 5 seconds.
The outer bath surfaces are resistant to treatment by wiping with disinfectants registered and approved in the Russian Federation for disinfection of surfaces according to the regimes regulated by the effective documents on the use of disinfectants approved in the established manner.
As for the methods of personnel protection against electric shock (electrical safety), the bath corresponds to class II type B as per GOST 50267.0-92.
    © LLC Ferroplast Medical, 2020