Ultrasound bath ВУ-12-"Я-ФП"-03

All parts of the bath (body, working container, cover) are made of stainless steel.
All parts of the bath (body, working container, cover) are made of stainless steel.

A piezoelectric element is installed at the bottom of the bath, which converts ultrasonic energy supplied from the generator into mechanical vibrations of the bath bottom and walls leading to cavitation in the washing solution.

The following controls are located on the front panel:

"Power" button

Digital treatment time indicator.

UV emitter operation indicator.

Heater operation indicator.

Digital treatment temperature indicator.

"+" and "-" buttons of treatment timer

"Start/Stop" button.

"+" and "-" buttons to adjust the liquid temperature.

The bath is equipped with a drain tap to drain the spent washing solution for convenience of the medical personnel.
MAIN PERFORMANCE VALUES

Ultrasonic transducer resonance frequency, kHz, - 40±2.

Temperature regulator, 0C. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . from room temperature to +60

Treatment procedure timer, min. . . . 1-99.

Supply voltage, V. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22010%, 50 Hz

The bath overall dimensions are 530х330х355 mm*
* Permissible deviation, no more than ± 5 mm.

The bath working container overall dimensions are 490х290х200 mm*
* Permissible deviation, no more than ± 5 mm.

The bath working container volume - 27.5 l*
* Permissible deviation, no more than  0.1 l.

The bath working container useful volume - 25 l*
* Permissible deviation, no more than  0.1 l.

Bath weight (outfitted) - 16.4 kg*
* Permissible deviation, no more than ± 0.5 kg.

Power consumed by the bath - no more than 900 VA

The liquid cavitation is provided in the bath over the entire useful volume of the working container. The cavitation effect is provided by the ultrasonic transducer with frequency of 40 kHz.

The bath is equipped with a treatment procedure timer providing for control of instruments treatment time.

Timer operation parameters:

Range 99

Increment 1
*Permissible deviation ± 0.4

A temperature regulator is provided in the bath to control the liquid temperature in the useful volume of the working container during the instruments treatment. Temperature regulator operation parameters, 0 С

Range 60

Increment 1
*Permissible deviation* ± 2
*- when using the ultrasonic cleaning mode with the interval of more than 30 minutes, the excess of preset temperature is permitted. This is due to the cavitation effect, during which additional heat is released. If long-time cleaning modes with maintaining certain temperature are required, it is recommended to open the cover for effective heat transfer.

The bath has a metal mesh located in the working container with the help of holders to accommodate the instruments withstanding uniformly-distributed load of 60 N

The time to recover the bath operation mode after switching on does not exceed 5 seconds.

The outer bath surfaces are resistant to treatment by wiping with disinfectants registered and approved in the Russian Federation for disinfection of surfaces according to the regimes regulated by the effective documents on the use of disinfectants approved in the established manner.

As for the methods of personnel protection against electric shock, the bath corresponds to class 1 type B as per GOST 50267.0-92.
© LLC Ferroplast Medical, 2020
Our Company in social networks
Contact us
Product
+